| Customization: | Available |
|---|---|
| Classification: | Imaging Diagnostic Equipment |
| Type: | X Ray Equipment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency



Part | Model | Technical parameters and Performance description |
High Voltage Generator | DTXR-50 | Power:32kw/45kw; Tube Current:10-500mA; Current time product: 0.1mAs~630mAs; Tube Voltage:40~150KVP; Can be directly connected to the network cable remote control for remote fault diagnosisand maintenance. |
X-ray tube(imported Toshiba original) | E7242X | Dual-focus 0.6mm/1.5mm, power 20/46KW, maximum current 500mA maximum voltage> 125KVP |
Amorphous siliconpanel Detector(lmportTFT Substrate 17*17) | CareView1800L | Strontium iodide amorphous silicon panel. Pixel matrix>2816X2816; lmaging area: 430x430mm" Pixelsize: <154um,Spatial resolution:3,5Lp/mm,AD conversion 16bits: MTF:> 60% @1D/mm>20%@3lp/mmDOE:(@ROA5,30uGy)>65% @0lp/mm >20% @3lp/mm Dynamic Range: >80dBlmaging time.The output time of the original image is less than 3 seconds, and the post-loading software processingshould be about 8 seconds. |
Flat bed | SYC-2 | Detector longitudinal motion range: 1000mmTable lateral movement 180 mm.Mechanical Motion Control: Manual electromagnetic lock.SlD1000 Korean Vibration Grid |
Tube stand | ZJ-1 | The X-ray tube assembly moves up and down no less than 900mm;Longitudinal movement is not less than 1800mm; |
Beam limiter | XSQ-1 | The total number of deviations between the sides of the ray radiation field and the respective sides ofthe light field does not exceed 3% of the distance from the field plane to the focal point. The sum of thedeviations of the two axes should not exceed 4% of the indicated focal point to the detector plane. |
Photograph Frame | SYJ-1 | Detector activity: Vertical movement no less than 900mm;SlD1000 Korean Vibration Grid |
DR image processing Software work station | Picke-DR | DR special diagnostic software:DlCOM3.0 standard, pre-selection of projection site, preset imagebrightness/contrast, window width/window position; organization equalization technology preset;image enhancement and sharpening technology preset; image segmentation, transmission, printing.etc. Can be directly connected to a remote consultation system |
Computer Host(DELL) | Picke2011 | CPU 2GHz or more, memory>4G hard disk> 1000G, can store pictures in large capacity. |
Display(DELL) | ML2002 | 24" professional display: maximum resolution: 2560x1440@75Hz; light transmittance: 37.5%;brightness:> 600 nits; scanning level range: 30-85kHz, automatic scanning vertical range: 50-120Hzautomatic scanning signal: BNC 15 pin D type; |
Remote Box | Can be remote controlled tube stand and photography frame, suitable for large-size chest X-ray examination | |
Documents | Registration document, manual, product certification | |